Xie Bin, Shi Yujie, Yi Guoqing, Wang Chao. Development of the 70kN strain gauge balance calibration system[J]. Journal of Experiments in Fluid Mechanics, 2014, (5): 71-75. DOI: 10.11729/syltlx20130078
Citation: Xie Bin, Shi Yujie, Yi Guoqing, Wang Chao. Development of the 70kN strain gauge balance calibration system[J]. Journal of Experiments in Fluid Mechanics, 2014, (5): 71-75. DOI: 10.11729/syltlx20130078

Development of the 70kN strain gauge balance calibration system

  • To meet the static calibration requirement of the balances used in force tests of 2 meter scale high speed wind tunnel,a 70kN strain gauge balance calibration system is developed. Based on the traditional OFAT method,the scheme for four degree of freedom,balance body ax-is system replacement and automatic calibration is adopted.Standard weights are applied to pro-vide accurate calibration loads by using automatic loading weights device,which controls weights independently.Through the double cross hinge force acting point positioning device and the but-terfly shape load-adapter,balance can be loaded exactly.Gapless replacement transmission mech-anism ensures precise replacement of the balance.The final calibration system can meet the cali-bration requirements of different types of balances such as sting and ring ranging from 20kN to 70kN loading.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return