高压交变流动下热线风速仪标定方法研究

论立勇, 陈厚磊, 蔡京辉

论立勇, 陈厚磊, 蔡京辉. 高压交变流动下热线风速仪标定方法研究[J]. 实验流体力学, 2010, 24(3): 87-91. DOI: 10.3969/j.issn.1672-9897.2010.03.018
引用本文: 论立勇, 陈厚磊, 蔡京辉. 高压交变流动下热线风速仪标定方法研究[J]. 实验流体力学, 2010, 24(3): 87-91. DOI: 10.3969/j.issn.1672-9897.2010.03.018
LUN Li-yong, CHEN Hou-lei, CAI Jing-hui. Investigation on calibration method of hot-wire anemometer in high pressure reciprocating flow[J]. Journal of Experiments in Fluid Mechanics, 2010, 24(3): 87-91. DOI: 10.3969/j.issn.1672-9897.2010.03.018
Citation: LUN Li-yong, CHEN Hou-lei, CAI Jing-hui. Investigation on calibration method of hot-wire anemometer in high pressure reciprocating flow[J]. Journal of Experiments in Fluid Mechanics, 2010, 24(3): 87-91. DOI: 10.3969/j.issn.1672-9897.2010.03.018

高压交变流动下热线风速仪标定方法研究

基金项目: 

国家自然科学基金重大项目

详细信息
  • 中图分类号: 11-5266/V

Investigation on calibration method of hot-wire anemometer in high pressure reciprocating flow

  • 摘要: 热线风速仪是重要的测速手段之一,但在高压交变流动条件下缺乏有效的标定方法.提出了一种通过测量气库内压力变化并按绝热热力学过程计算进出气库气体流速从而标定热线风速仪的方法.为了对该标定方案的误差进行分析,建立了该标定方案的数值模型,运用CFD软件FLUENT对该方案气库壁面分别设为绝热和等温边界条件时的热力过程进行了数值模拟,结果表明标定方案中采用绝热物理模型所引起的误差最大不超过2%.由于在热线探针两侧布置了一定量丝网作为均流元件,热线探针所测速度与管路截面平均速度误差不超过1.4%.该标定方法总误差不超过3.37%.
计量
  • 文章访问数:  112
  • HTML全文浏览量:  23
  • PDF下载量:  9
  • 被引次数: 0
出版历程
  • 刊出日期:  2010-02-28