电弧加热射流O2激光诱导荧光技术温度测量

岳斌, 苏铁, 余福明, 杨富荣, 张传斌

岳斌, 苏铁, 余福明, 杨富荣, 张传斌. 电弧加热射流O2激光诱导荧光技术温度测量[J]. 实验流体力学, 2004, 18(1): 84-88. DOI: 10.3969/j.issn.1672-9897.2004.01.019
引用本文: 岳斌, 苏铁, 余福明, 杨富荣, 张传斌. 电弧加热射流O2激光诱导荧光技术温度测量[J]. 实验流体力学, 2004, 18(1): 84-88. DOI: 10.3969/j.issn.1672-9897.2004.01.019
LIF temperature measurement of arcjet by O2-laser induced fluorescence[J]. Journal of Experiments in Fluid Mechanics, 2004, 18(1): 84-88. DOI: 10.3969/j.issn.1672-9897.2004.01.019
Citation: LIF temperature measurement of arcjet by O2-laser induced fluorescence[J]. Journal of Experiments in Fluid Mechanics, 2004, 18(1): 84-88. DOI: 10.3969/j.issn.1672-9897.2004.01.019

电弧加热射流O2激光诱导荧光技术温度测量

详细信息
  • 中图分类号: 51-1499/V

LIF temperature measurement of arcjet by O2-laser induced fluorescence

  • 摘要: 采用ArF准分子激光器作为激发光源来激励O2-Schumann-Runge带系(B3Σ-u←X3Σ-g)193.3nm附近的转动吸收谱线,用紫外光电倍增管探测该带系256nm附近的荧光信号,建立了ArF准分子激光器O2激光诱导荧光流场测试系统.在建立O2单带测温模型和软件的基础上,利用Schumann-Runge带系(10,2)带的荧光信号、测量了氧氩混合电弧加热射流的径向温度分布.结果表明:我们建立的ArF准分子激光器O2-LIF测试系统是成功的.